TD-4100 Oil in Water
Monitor
The Turner Designs TD-4100 reliably detects BTEX, gasoline, diesel, jet
fuel and oil in water. The TD-4100 is a nonfouling continuous, on-line monitor
that provides hydrocarbon and oil in water monitoring down to low ppb. The
TD-4100 is installed for applications that require monitoring of treated and
untreated wastewater streams ranging from industrial wastewater, groundwater,
process water, produced, potable water intakes, and storm water
run-off.
Why Continuous On-Line Monitoring?
Continuous on-line monitoring provides the most responsive feedback loop
for measuring hydrocarbons in water. Continuous monitoring is reliable,
effective, and recognized for its ability to improve process management for
treating, discharging and detecting hydrocarbons in water. Compared to
laboratory grab sample analysis, on-line monitoring provides cost effective,
continuous, remote, operator unattended measurement of aromatic hydrocarbons in
water.
Features of the TD-4100
The Unique Flowcell. The TD-4100 does not have a
glass flow cell. Hydrocarbons are detected in a stream of water which falls
through an open chamber; the stream does not contact, dirty or foul the optical
windows used for oil in water monitoring. A proprietary Air Curtain system keeps
optical Windows fog-free in humid environments or hot water
applications.
Low Maintenance. There are no tubes, pumps or valves to replace.
The system’s one moving part has a life expectancy of greater than 5 years.
Routine maintenance involves changing a lamp twice a year and an air filter
periodically.
Direct, Continuous Monitoring. The TD-4100 monitors a flowing
water stream continuously. No chemicals, no pre-treatment and no manual
manipulation of the sample are required to monitor aromatic
hydrocarbons.
Accurate. The TD-4100 directly measures aromatic hydrocarbons in
water with accuracies that consistently correlate to regulated laboratory
methods.
Sensitive. BTEX, gasoline, diesel, jet fuel, crude oil, aromatic
solvents and refined petroleum products containing aromatics are detected by the
TD-4100 from low ppb (ug/L) to high ppm (mg/L). For example, the TD-4100 can
detect 5 ppb of gasoline or diesel fuel in water free of interfering
compounds.
Selective. The TD-4100 continuously measures the fluorescence of
aromatic hydrocarbon compounds in water. Fluorescence occurs when a molecule
absorbs light energy of one specific wavelength and emits light energy of a
longer wavelength. Fluorescent compounds each have a unique signature, and these
compounds can be displayed as an actual concentration.
Effective Monitoring in Dirty Water. Monitoring fluorescence makes
the TD-4100 resistant to interferences by turbid or dirty water which impact
on-line UV or IR light scatter instruments. Most substances absorb light, but
very few fluoresce; if a substance does not fluoresce at specific wavelengths
for the monitored aromatic hydrocarbon, it will not
interfere.
Operator Friendly. The TD-4100 is designed for easy operation.
Simple on-board software controls alarms, 4-20 mA output, diagnostics and
calibration.
|
Applications exist in the fields of...
- Upset Monitoring
- Discharge Monitoring
- Continuous Monitoring
- Process Control
- Carbon Break-thru
- Oil/Water Separators
- Treatment Verification
- Hydrocarbon Leak Detection
- Pollution Prevention
- Regulatory Compliance
|
 |
TD-4100 Specifications
| Dimensions: |
44.0" x 43.1" x 10.8" (111.8 cm x 109.4 cm x 27.5
cm) |
| Weight: |
64 lbs (29 kg) |
| Power Requirements: |
110 - 240 VAC, 50/60 Hz, 2A |
| Plumbing Requirements: |
Feed: 0.5 - 2 gpm (2 - 8L/min), 2" (5.01 cm)
unrestricted drain |
| Air Supply: |
Dry Air: 10 - 100 psi (68.9 - 689 kPa), 15 SCFH (246
cm3H) |
| Operating Temperature: |
32 - 120o F (0 - 49o
C). Contact Turner Designs for configurations in extreme
environmental conditions. |
| Operational Principle: |
Fluorescence |
| Detection Range: |
1 ppb - > 1000 ppm* |
| Calibration Stability: |
10% or better over 90 days |
| Response Time: |
< 10 sec |
| Sample Temperature: |
190o F (95o C) (max) |
| Calibration: |
Multi-point calibration |
| Alarms: |
Three alarm trigger points (baseline, early warning,
and high alarm). Critical components linked to alarms:
Local Alarm: Local beep tone; LCD alarm
message |
| Outputs: |
Signal: 4 - 20 mA, isolated. External
Alarm/Device Control: 2 user-settable,
independently-fused, solid-state relays (optional AC relay or
optional DC relay); relays configurable as Normally Open or
Normally Closed; user settable set points and
delays |
| Diagnostics: |
Self diagnostics: Internal failure mode linked to
relays and LCD display |
| Security: |
Password protected for Master and User ID; padlockable
cabinet |
| Electronics Cabinet: |
NEMA 4X |
| Maintenance: |
Lamp (twice yearly); routine cleaning of
plumbing |
| Air Purge (optional): |
Class I Division I; Class I Division II
(explosion-proofing) |
| Data Logger (optional): |
Digital data
storage |
|